Evident LogoOlympus Logo

Product Information

21 March, 2024

Important Notice Regarding Colony Count & Size Analysis of Incubation Monitoring Systems CM30 and CM20

Dear Valued Evident Customer,

Thank you for your continued business with Evident Corporation.
We have found a problem with our incubation monitoring systems CM30 and CM20 in which the results of the Colony Count & Size analysis differ from the actual measured values.

The affected products and software are as follows. We apologize for the inconvenience and ask that you please apply the Service Update and batch files.

<Affected products and software> 

  • Incubation Monitoring System CM30 Software Version 2.2.1 or earlier
  • OLYMPUS Provi CM20 Software Version 1.2.6 or earlier

<Details of the problem>

When executing the "Colony Count & Size Analysis" on the incubation monitoring systems CM30 and CM20, if the project is executed with the Cell Type set to "Colony Forming", the colony size displayed in the "Colony Count & Size Graph" is different from the actual measured value in the monitoring results.

Figure: Image of the problems

Figure: Image of the problems

<Improvement methods>

These problems can be resolved by applying Service Update (CM30 Software Version 2.2.2, CM20 Software Version 1.2.7). However, since the Service Update is not automatically applied to data that has already been analyzed in past projects, it is necessary to re-analyze past projects using the batch file that will be separately prepared for you. 

For information on how to apply the Service Update and batch files, please refer to the following URL for details on the affected products and each file.

Download software update installer
Download batch file for Colony Count & Size reanalysis 

If you have any questions, please contact our sales representatives or customer service center.

Thank you for your understanding and cooperation.
 

Back to Product Information

Sorry, this page is not
available in your country.

Sorry, this page is not available in your country