Main Laser Combiner | 405 nm: 50 mW, 488 nm: 100 mW, 561 nm: 100 mW, 640 nm: 100 mW | |||
---|---|---|---|---|
Laser Light Control | Main laser combiner with implemented AOTF system, ultra-fast intensity modulation with individual laser lines, additional shutter control continuously variable (0.1 % - 100 %, 0.1 % increments) | |||
Scanner | Yokogawa CSU-W1 |
Single 50 μm pinhole disk
Optionally available as a dual disk system with selectable 50 μm and 25 μm disks | ||
Super Resolution imaging | Sensor | Single camera detection for realtime single channel super resolution detection | Hamamatsu Orca Flash 4 V2+ High Speed SCMOS camera | |
Frame Rate (Max) | 3 fps | |||
Optical Zoom | 3.2 X | |||
Optical Resolution | 120 nm | |||
Pinhole | 50 μm | |||
Field Number | 5.6 | |||
Standard Resolution Imaging | Sensor | Optional dual camera model for simultaneous dual channel monochrome imaging | Hamamatsu Orca Flash 4 V2+ High Speed SCMOS camera | |
Frame Rate | 200 fps | |||
Optical Zoom | 1 X | |||
Optical Resolution | 190 nm | |||
Pinhole |
50 μm (single disk)
Optional motorized dual disk model includes 25 μm disk, WF bypass mode | |||
Field Number | 18 | |||
Dichromatic Mirror | 3 position (motorized slider) | |||
Filter Wheel (emission) | 10 position (motorized wheel) | |||
Microscope | Motorized Microscope | Inverted IX83 (IX83P2ZF) | ||
Motorized Stage | IX3-SSU | |||
Super Resolution Adapter | 1 X / 3.2 X (Manual change) | |||
Workstation | PC |
OS: Windows 7 Professional 64-bit (English version)
Xeon Processor E5-1650 (12MB, 3.2 GHz, 6 core or more) 1st HDD: 512 GB SSD, 2nd HDD (1TB), 3rd HDD (1TB) | ||
Display | 30 or 32-inch monitor (WQUXGA 2560 x 1600) | |||
Imaging software | MetamMorph Advanced for Olympus | Multi-Dimensional Acquisition and analyiss | ||
SD-OSR Super Resolution Imaging Module w/ raw data export |