Laser Unit > Qualified IR Pulsed Laser with Negative Chirp for Multiphoton Excitation

-

Laser Unit > Automatic Introduction Optic

Built-in
Continuously variable 
(0.1 % - 100 %, 0.1 % increment)
REX: Capable of laser intensity adjustment and laser wavelength selection for each region.

Laser Unit > Optional Visible Light Laser for Stimulation

-

Laser Unit > LD Laser for Visible Light > 405 nm
  • 17,1 mW
Laser Unit > LD Laser for Visible Light > 440 nm
  • -
Laser Unit > LD Laser for Visible Light > 473 nm
  • 11,9 mW
Laser Unit > LD Laser for Visible Light > 559 nm
  • 15 mW
Laser Unit > LD Laser for Visible Light > 635 nm
  • 20 mW
Laser Unit > HeNe(G) Laser (543 nm, 1 mW)
  • -
Scanning and Detection > Main Scanner > Standard Laser Ports
  • 1 (incorporado)
Scanning and Detection > Main Scanner > Detector > Standard
  • 2
Scanning and Detection > Main Scanner > Detector > Cooled GaAsP-PMT 2 CH
  • -
Scanning and Detection > Main Scanner > Detector > Optional 4 CH
  • -
Scanning and Detection > Main Scanner > Photo Detection Method > Analog Integration
Scanning and Detection > Main Scanner > Photo Detection Method > Hybrid Photon Counting
  • -
Scanning and Detection > Main Scanner > VIS - UV - IR Excitation Dichromatic Mirror Turret
  • 1 posición (fija)
Scanning and Detection > Main Scanner > Beamsplitter Turrets
  • -
Scanning and Detection > Main Scanner > Wavelength Selection

Variable barrier fi lter mechanism for fl uorescence channel by diffraction grating and slit

Scanning and Detection > Galvanometer Scanner (Normal Imaging) > Galvanometer Mirror Scanner (X, Y)
  • Escáner doble
Scanning and Detection > Galvanometer Scanner (Normal Imaging) > Scanning Modes > 2D

XY

Scanning and Detection > Galvanometer Scanner (Normal Imaging) > Scanning Modes > 3D

XYT, XYZ

Scanning and Detection > Galvanometer Scanner (Normal Imaging) > Scanning Modes > 4D

XYZT

Scanning and Detection > Galvanometer Scanner (Normal Imaging) > Scanning Modes > 5D

-

Scanning and Detection > Galvanometer Scanner (Normal Imaging) > Scanning Modes > Other

Map image, one shot
Rotation scanning: 0-359.9° in 0.1° increments

Scanning and Detection > Galvanometer Scanner (Normal Imaging) > Scanning Speed
  • 512 x 512 (1,1 s, a máxima velocidad)
Scanning and Detection > Galvanometer Scanner (Normal Imaging) > Pinhole

Single motorized
Diameter ø50 – 800 μm 
automatic setting (adjustable to x1.0, x1.5, x2.0, x2.5, x3.0, x4.0, and x5.0)

Scanning and Detection > Galvanometer Scanner (Normal Imaging) > Scanning Zoom
  • 10 X: 1 X - 6 X en incrementos de 0,1 X
  • 60 X: 1 X - 10 X en incrementos de 0,1 X
Scanning and Detection > Resonant Scanner (High-Speed Imaging) > Scanning Modes > 2D

-

Scanning and Detection > Resonant Scanner (High-Speed Imaging) > Scanning Modes > 3D

-

Scanning and Detection > Resonant Scanner (High-Speed Imaging) > Scanning Modes > 4D

-

Scanning and Detection > Resonant Scanner (High-Speed Imaging) > Scanning Modes > 5D

-

Scanning and Detection > Resonant Scanner (High-Speed Imaging) > Scanning Modes > other

-

Scanning and Detection > Resonant Scanner (High-Speed Imaging) > Scanning Speed

-

Scanning and Detection > Resonant Scanner (High-Speed Imaging) > Scanning Zoom

-

Scanning and Detection > Field Number (NA)
  • 18
Scanning and Detection > Z-Drive
  • Enfoque motorizado incorporado
  • Incremento mínimo: 0,01 μm
Scanning and Detection > Transmitted Light Detector Unit

Phase Contrast: 1 channel

Microscope > Frame
  • -
Microscope > Frame > Inverted
  • Incorporado
Microscope > Objectives and Focus > Objectives
  • Objetivo 10 X de contraste de fase, diseño exclusivo / AN 0,4 (equivalente al UPLSAPO 10 X)
  • Objetivo 60 X de contraste de fase para inmersión en agua, diseño exclusivo / AN 1,2 (equivalente al UPLSAPO 60X W) / con collar de corrección motorizado
  • Sistema remoto de cambio desde el software por revólver eléctrico
Microscope > Objectives and Focus > Auto Focus (AF)
  • Detección automática de la interfaz entre la muestra y el cubreobjetos mediante detección de la reflexión del láser
  • Detección automática del grosor del cubreobjetos y ajuste automático del collar de corrección motorizado
Microscope > Objectives and Focus > Water Supply
  • Suministro automático de agua y mecanismo automático de limpieza por aire para el objetivo 60 X de inmersión en agua
Microscope > Objectives and Focus > Oil Supply
  • -
Microscope > XY Stage > XY Driving Method
  • Módulo de platina XY motorizada por motor de pasos
  • Incremento mínimo: 0,3 μm
Microscope > XY Stage > Specimen Holder
  • Sólo puede montarse el portamuestras específico
  • Para una placa de 35 mm de diámetro con fondo de vidrio
  • Para un portaobjetos de vidrio
  • Para una cámara cubreobjetos (de 8 pocillos)
  • Para portaobjetos de pocillos (de 8 pocillos)
  • Cápsula de cultivo (para una placa de 35 mm de diámetro con fondo de vidrio)
Microscope > Incubator > Room Environment
  • Temperatura: 37 + 0,1 °C, -0,5 °C (puede desactivarse)
  • Humedad: superior al 90%
  • Concentraciones de CO2: 5% (recomendada), 1 - conector de junta (ø2 mm) para el regulador de CO2 exterior
Microscope > Incubator > Heating Method
  • Calefacción sin contacto mediante calentador a resistencia montado en una sección del estativo
System Control > Controller
  • SO: Windows 7 Professional de 64 bits (versión en inglés)
System Control > Power Supply Unit

-

Optional Unit > SIM Scanner
  • -
Optional Unit > TIRFM Unit
  • -
Software > Image Acquisition

-

Software > Programmable Scan Controller

-

Software > 2D Image Display

-

Software > 3D Visualization and Observation

-

Software > Image Format

-

Software > Spectral Unmixing

-

Software > Statistical Processing

-

Software > Optional Software

-

Software > Image Acquisition Mode

Map image, one shot, time-lapse (XYT), Z-stack (XYZ), Z-stack time-lapse (XYZT), multi area time-lapse (Multi Area XYT), multi area 
Z-stack time-lapse (Multi Area XYZT)

Software > Map Image Acquisition

Automatic selection of map image of 3 x 3 – 35 x 7 fields according to 10 X objective lens (The maximum area varies in accordance to the specimen holder used), and manual selection of map acquisition area

Software > Multi Area Time-Lapse

Automatic multi area time-lapse by motorized XY stage 
Setting for each registered point: Image size, scanning speed, cross talk reduction, pinhole diameter, 
rotation angle, galvano zoom, acquisition channel, laser power, PMT sensitivity, Z condition 
Maximum resister number: 10 items per one container 
Maximum interval time: one hour 
Maximum acquisition number of times: 3000 times per one point 
software Z Drift Compentation (Only FV10i-LIV 60 X objectives) 

Software > Image Acquisition Area

Area appointment: All area, clipping square area (minimum area: 96 x 96 pixels)

Software > Image Display
  • Visualización por canal
  • Visualización superpuesta
  • Revisión de imagen en progreso
Software > Cross Talk Reduction

Line sequential action (2 channel), or frame sequential action (3 channel and 4 channel)

Software > Acquisition Image File Type
  • Formato de imagen OLYMPUS (OIF)
Software > Image File Type Available for Viewing
  • BMP
  • JPEG
  • Formato multi-TIFF (escala de grises de 8/16 bits, color Index, color de 24/32/48 bits)
  • Formato de imagen OLYMPUS (OIF, OIB)
  • TIFF
Software > Image Editing

LUT: pseudo color setting, contrast adjustment, Comment: inputting graphic, text, scale etc., image extraction, combination

Software > 3D Image Construction

3D display: AlphaBrend method, Maximum intensity projection method 3D animation display, free orientation of cross section display

Software > IR Laser Controlling
  • -
Dimensions, Weight and Power Consumption > Microscope with Scan Unit > Dimensions (mm)
  • 470 (A) x 680 (P) x 505 (Al.)
Dimensions, Weight and Power Consumption > Microscope with Scan Unit > Weight (kg)
  • 73
Dimensions, Weight and Power Consumption > Microscope with Scan Unit > Power Consumption
  • (Alimentado por FV10C-PSU)
Dimensions, Weight and Power Consumption > Fluorescence Illumination Unit > Dimensions (mm)
  • -
Dimensions, Weight and Power Consumption > Fluorescence Illumination Unit > Weight (kg)
  • -
Dimensions, Weight and Power Consumption > Fluorescence Illumination Unit > Power Consumption
  • -
Dimensions, Weight and Power Consumption > Transmitted Light Detection Unit > Dimensions (mm)
  • -
Dimensions, Weight and Power Consumption > Transmitted Light Detection Unit > Weight (kg)
  • -
Dimensions, Weight and Power Consumption > Transmitted Light Detection Unit > Power Consumption
  • -
Dimensions, Weight and Power Consumption > Microscope Control Unit > Dimensions (mm)
  • -
Dimensions, Weight and Power Consumption > Microscope Control Unit > Weight (kg)
  • -
Dimensions, Weight and Power Consumption > Microscope Control Unit > Power Consumption
  • -
Dimensions, Weight and Power Consumption > FV Power Supply Unit > Dimensions (mm)
  • 230 (A) x 330 (P) x 150 (Al.)
Dimensions, Weight and Power Consumption > FV Power Supply Unit > Weight (kg)
  • 7,5
Dimensions, Weight and Power Consumption > FV Power Supply Unit > Power Consumption
  • CA 100-120/200-240 V 50/60 Hz 5,0 A/2,5 A
Dimensions, Weight and Power Consumption > Power Supply Unit for Laser Combiner > Dimensions (mm)
  • 210 (A) x 300 (P) x 100 (Al.)
Dimensions, Weight and Power Consumption > Power Supply Unit for Laser Combiner > Weight (kg)
  • 4
Dimensions, Weight and Power Consumption > Power Supply Unit for Laser Combiner > Power Consumption
  • CA 100-120/220-240 V 50/60 Hz 2,0 A/1,0 A
Dimensions, Weight and Power Consumption > Laser Combiner (with Ar Laser Heads) > Dimensions (mm)
  • 514 (A) x 504 (P) x 236 (Al.)
Dimensions, Weight and Power Consumption > Laser Combiner (with Ar Laser Heads) > Weight (kg)
  • 40
Dimensions, Weight and Power Consumption > Laser Combiner (with Ar Laser Heads) > Power Consumption
  • -
Dimensions, Weight and Power Consumption > Laser Combiner (without Ar Laser Heads) > Dimensions (mm)
  • 514 (A) x 364 (P) x 236 (Al.)
Dimensions, Weight and Power Consumption > Laser Combiner (without Ar Laser Heads) > Weight (kg)
  • 40
Dimensions, Weight and Power Consumption > Laser Combiner (without Ar Laser Heads) > Power Consumption
  • -
Dimensions, Weight and Power Consumption > LD559 Laser Power Supply > Dimensions (mm)
  • 200 (A) x 330 (P) x 55 (Al.)
Dimensions, Weight and Power Consumption > LD559 Laser Power Supply > Weight (kg)
  • 1,2
Dimensions, Weight and Power Consumption > LD559 Laser Power Supply > Power Consumption
  • CA 100-240 V 50/60 Hz 30 W
Dimensions, Weight and Power Consumption > Multi Ar Laser Power Supply > Dimensions (mm)
  • 162 (A) x 287 (P) x 91 (Al.)
Dimensions, Weight and Power Consumption > Multi Ar Laser Power Supply > Weight (kg)
  • 4,4
Dimensions, Weight and Power Consumption > Multi Ar Laser Power Supply > Power Consumption
  • CA 100-240 V 50/60 Hz 20 A
Dimensions, Weight and Power Consumption > HeNe(G) Laser Power Supply > Dimensions (mm)
  • 130 (A) x 224 (P) x 65 (Al.)
Dimensions, Weight and Power Consumption > HeNe(G) Laser Power Supply > Weight (kg)
  • 1,8
Dimensions, Weight and Power Consumption > HeNe(G) Laser Power Supply > Power Consumption
  • CA 200-240 V 50/60 Hz 0,23 A
Dimensions, Weight and Power Consumption > Operating Environment (Indoor Use) > Ambient Temperature18 - 28 ℃ (fluctuation ±2 ℃)
Dimensions, Weight and Power Consumption > Operating Environment (Indoor Use) > Maximum Relative Humidity30 - 80% (sin condensación)