3D Time-Lapse Imaging of Spheroids with the FV3000 Confocal Microscope: 48-Hour Continuous Observation of Antibody-Dependent Cell-Mediated Cytotoxicity (ADCC)


Antibody-dependent cell-mediated cytotoxicity (ADCC) is one of the mechanisms of action of antibody drugs. Effector cells, such as natural killer (NK) cells and monocytes, recognize target cells (such as cancer cells) that have been bound by an antibody. Upon recognizing the antibody, the effector cells then destroy the target cells by releasing cytotoxic factors that induce cell lysis. In this application, we performed a three-dimensional co-culture experiment using human colon cancer cells (HT-29) and NK cells (KHYG-1) expressing a functional Fcgamma (FCγ) receptor in the presence of human EGFR-targeted therapeutic monoclonal antibodies, and observed them for forty-eight hours using the FLUOVIEW FV3000 confocal microscope. The resulting three-dimensional (3D) imaging of ADCC enabled us to capture how NK cells attack cancer cells.

3D Live-Cell Spheroid Imaging with Reduced Phototoxicity

Cell masses grown in a 3D culture system (called spheroids) are difficult to image due to the thickness of the mass. Higher excitation light intensity is typically required to produce a fluorescent signal strong enough to be detected. However, higher excitation light also increases phototoxicity, resulting in damage to the cells. To combat this, we used the FV3000 microscope, which incorporates Olympus’ TruSpectral detection technology to capture weak fluorescent signals and minimize the laser power required for signal generation deep within a spheroid. Furthermore, we used a 60X silicone oil immersion objective (closely matching the refractive index of live cells) to achieve brighter images with higher resolution for deep observation. By combining these two technologies, we successfully captured bright 3D time-lapse images of a spheroid for 48 hours, as it collapsed due to NK cell-mediated HT-29 cell death.


Fig. 1: Movie obtained by 3D imaging NK cells attacking spheroid HT-29 tumor cells for 48 hours

NK cell line KHYG-1 (green) expressing functional Fcgamma (FCγ) receptors and ZsGreen fluorescent proteins were cultured with spheroids of HT-29 cells (cell lines derived from colorectal cancer). The surfaces of HT-29 cells were labeled with anti-EGFR monoclonal antibody cetuximab (Alexa Fluor 647, blue). HT-29 and NK cells were cultivated in the presence of propidium iodide (PI), and XYZT imaging was performed over the course of 48 hours. Nuclei labeled red (marking cellular uptake of PI) was used as an indicator of cell death.

Imaging conditions
Objective: 60X Silicone oil immersion objective (UPLSAPO60XS)
Microscope: FLUOVIEW FV3000 System
Lasers: 488 nm (ZsGreen, green), 561 nm (PI, red), 640 nm (Alexa Fluor 647, blue)

Z-Drift Compensation System Enables Accurate Time-Lapse Imaging

For experiments that evaluate the effectiveness of antibody drugs, long-term imaging of live cells is indispensable. To more precisely capture the effect of NK cells on spheroidal cancer cells,  we used a Z-drift compensator (IX3-ZDC2) to prevent changes in focus due to external environmental changes such as temperature variation. The IX3-ZDC2 enabled us to obtain a time-lapse image over a long period without losing focus on the spheroid. This enabled us to confirm that the NK cells actively change their shape while attacking and entering the tumor mass.

9h
9h
11h
11h
17h
17h
22h
22h
24h
24h

Fig. 2: NK cell line KHYG-1 (green) changing shape while attacking and killing HT-29 tumor cells labeled with cetuximab (blue). PI uptake (red) indicates cell death.

How the FV3000 Confocal Microscope Facilitated Our Experiment

Fully Spectral System Provides High Sensitivity

The FV3000 series employs Olympus' TruSpectral detection technology that diffracts light via transmission through a volume phase hologram unit. This technology enables much higher light throughput when compared to conventional spectral detection units with reflection-type gratings, and minimizes the laser power needed for deep tissue observation.

Maintain Focus with Z-Drift Compensation (ZDC) System

The IX3-ZDC2 Z-drift compensator uses minimally-phototoxic infrared light (laser class 1) to identify the location of the sample plane. One-shot autofocus (AF) mode enables several focus positions to be set as desired for deeper samples, enabling efficient Z-stack acquisitions in multiposition experiments.

Silicone Immersion Objectives for Live-Cell Imaging Deliver High-Resolution Observation at Depth

The refractive index of silicone oil (ne≈1.40) is close to that of living tissue (ne≈1.38), enabling high-resolution observations deep inside living tissue with minimal spherical aberration caused by refractive index mismatch. Silicone oil also does not dry out or harden, so there is no need to refill the oil, making it ideal for extended time-lapse observations.

Comment by Dr. Yuji Mishima

Dr. Yuji Mishima

Dr. Yuji Mishima

In recent years, the demand for fluorescence imaging in cancer research has been increasing, as have the number of experiments required to sterically evaluate the effectiveness of antibody medicine—especially with spheroid cells. Spheroids are very difficult to observe using fluorescence due to their thickness, but by using the FV3000 microscope, we were able to observe spheroid cancer cells throughout their entirety for 48 hours. Throughout the ADCC experiment, we were able to accurately measure living cancer cells, dead cells, and effector cells (NK cells) as they responded to the antibody treatment. I felt the FV3000 was very effective for evaluating the therapeutic effectiveness of antibody medicine.

Acknowledgments
This application note was prepared with the help of the following researchers: 
The Cancer Chemotherapy Center of JFCR, Dr. Yuji Mishima

Products used for this application

激光扫描共聚焦显微镜

FV3000

  • 所提供的两种配置为仅装有常规振镜扫描单元的FV3000显微镜和装有常规/共振混合
  • 型扫描单元的FV3000RS
  • 可对所有通道进行全新、高效、精确的TruSpectral全真光谱检测
  • 经过优化,可以完成具有高灵敏度和低光毒性的活细胞成像操作
超级复消色差物镜

UPLSAPO

  • 为从可见光到近红外区域的球差和色差提供全补偿。
  • 在荧光、明场和Nomarski DIC观察时获取锐利、清晰的图像,没有色彩偏移。
  • 为各种数码成像的质量和性能需求提供无与伦比的解决方案 – 包括在活体组织内使用硅油物镜实施的高分辨率观察。
Z drift compensator

IX3-ZDC2

  • 随时保持焦点
  • 易用性设计
  • 专为活细胞影像设计
  • 通过cellSens软件实现高精度,多区域成像

已经成功添加到您的书签

查看书签关闭

Maximum Compare Limit of 5 Items

Please adjust your selection to be no more than 5 items to compare at once

对不起,此内容在您的国家不适用。

This site uses cookies to enhance performance, analyze traffic, and for ads measurement purposes. If you do not change your web settings, cookies will continue to be used on this website. To learn more about how we use cookies on this website, and how you can restrict our use of cookies, please review our Cookie Policy.

OK